Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 30: CAL ADJ /5520 DATE: 18-Feb-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 96 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # Fluke Model 30 Clamp Meter Service Information (PN 929042 January 1994). # Fluke 5520A/COIL Instruction Sheet (PN 600616 Aug. 1995) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- W 1.003 HEAD EQUIPMENT SETUP 1.004 DISP This procedure requires a 50-turn Toroid Coil. 1.004 DISP Refer to the 31/33 Instruction Manual figure 7, page 32 1.004 DISP for directions for constructioning a suitable coil. 1.005 HEAD CALIBRATION ADJUSTMENT 1.006 DISP Disassemble the meter by removing the four screws on the 1.006 DISP case back. 1.007 DISP Remove the battery cover and front panel. 1.008 DISP Turn on the meter and use a DMM to verify that the 1.008 DISP battery level is between 7 and 9 volts. If the battery 1.008 DISP does not meet this criteria, replace the battery before 1.008 DISP proceeding with the procedure. 1.009 DISP The Test equipment should be allowed to warmup and 1.009 DISP stablize for at least 30 minutes before starting the 1.009 DISP adjustments. 1.010 DISP Connect the 5520A NORMAL Hi & Lo to the UUT V Ohm & COM 1.010 DISP terminals. 1.010 DISP 1.010 DISP Connect the 5500A/COIL to the 5520A 20A & AUX Lo 1.010 DISP terminals. 1.010 DISP 1.010 DISP Clamp the meter around the toroid center conductor 1.010 DISP bundle, positioning at the alignment marks. 1.011 HEAD CALIBRATION ADJUSTMENT: AC Voltage 1.012 RNG 200V AC 1.013 5520 100.0V 60H SI S 2W 1.014 DISP Adjust VR3 for a UUT reading of 100.0V. 1.015 5520 * S 1.016 HEAD CALIBRATION ADJUSTMENT: AC Current 1.017 RNG 200A AC 1.018 5520 191.0A 60H SI HC S LT5 1.019 DISP Adjust VR2 for a UUT reading of 191.0A. 1.020 5520 * S 1.021 RNG 400A AC 1.022 5520 350A 50H SI HC S LT5 1.023 MEMI Enter UUT reading in amps: 1.024 MATH MEM1 = MEM - 350 1.025 JMPZ 1.028 8U 1.026 DISP Balance the calibration between the 200A and 400A ranges 1.026 DISP so that both ranges fall within tolerance. 1.027 JMP 1.016 1.028 DISP Remove the UUT from the toroid. 1.028 DISP Disconnect the toroid from the 5520A. 1.029 HEAD CALIBRATION ADJUSTMENT: Continuity Beeper 1.030 RNG 200Z 1.031 5520 30.1Z S 2W 1.032 DISP Adjust VR4 until the beeper sounds. 1.033 5520 * S 1.034 DISP Disconnect the 5520A from the UUT. 1.035 HEAD CALIBRATION ADJUSTMENT: Low Battery Indication 1.036 DISP Disconnect the battery from the UUT. 1.037 DISP Connect the 5520A Output to the UUT battery terminals. 1.038 5520 6.9V S 2W 1.039 DISP Adjust VR1 until the low battery indictor 1.040 5520 7.0V S 2W 1.041 OPBR Is the low battery indicator off? 1.042 JMPF 1.035 1.043 END